8 - 7
PM
code Description
What to do/What to check
D
Emergency Lowering Device
Verify functionality of the
emergency lowering device
NOTE
THE EMERGENCY
LOWERING DEVICE CAN
BE VERIFIED WITH OR
WITHOUT ATTACHING A
LOAD.
1. Pull the red emergency cord to turn off power.
2. Open the small side door to access the lowering mechanism.
3. Remove the 8 mm hex key on the top of the ceiling lift. Insert the hex key deep
into the socket.
4. Turn the hex key counter-clockwise to slowly lower the spreader bar.
WARNING
DO NOT USE THE UP/DOWN FUNCTIONS WHEN THE HEX KEY IS
INSERTED INTO THE SOCKET AS IS CAN LEAD TO INJURIES.
Auxiliary Up/Down Button
Verify functionality of the
alternate Up/Down buttons
1. Press and hold each button (Up and Down).
WITH THE DOWN FUNCTION, ALWAYS KEEP TENSION (AT LEAST
THE WEIGHT OF THE 2 HOOKS SPREADER BAR) ON THE STRAP.
THIS WILL PREVENT THE DOWN FUNCTION FROM STOPPING.
NOTE
001-15697 rev
. 5
Summary of Contents for Maxi Sky 2
Page 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Page 2: ...NOTE This page is left blank intentionally ...
Page 6: ...NOTE This page is left blank intentionally ...
Page 8: ...NOTE This page is left blank intentionally ...
Page 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Page 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Page 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Page 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Page 133: ...8 20 001 15697 rev 5 ...
Page 134: ...8 21 001 15697 rev 5 ...
Page 135: ...8 22 001 15697 rev 5 ...
Page 137: ...8 24 001 15697 rev 5 ...
Page 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...