8 - 16
For software version 2.2.X and up
There are 2 ways to reset a maintenance code
1. When maintenance has been performed, follow the steps described on page 8-15 to access the maintenance
menu.
- Press
button to select ‘‘maintenance required’’.
- Confirm you selection by pressing on the button.
- Select a maintenance letter that has been performed by pressing on the
button.
- Select ‘‘yes’ and press
button, then the letter associated to this maintenance will disappear from the list.
- Repeat the same sequence for each maintenance action performed.
2. To reset the letter of a maintenance performed before it is due and pops up on the screen, follow the steps described on
page 8-15 to access the maintenance menu.
- Press
button to select ‘‘preventive maintenance scheduled’’.
- Confirm your selection by pressing on the
button
- Select a maintenance letter to be reset by pressing on the
button
- Select ‘‘yes’’ and press
button, then the letter associated to this maintenance will disappear from the list.
001-15697 rev
. 5
Summary of Contents for Maxi Sky 2
Page 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Page 2: ...NOTE This page is left blank intentionally ...
Page 6: ...NOTE This page is left blank intentionally ...
Page 8: ...NOTE This page is left blank intentionally ...
Page 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Page 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Page 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Page 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Page 133: ...8 20 001 15697 rev 5 ...
Page 134: ...8 21 001 15697 rev 5 ...
Page 135: ...8 22 001 15697 rev 5 ...
Page 137: ...8 24 001 15697 rev 5 ...
Page 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...