001-31249-EN rev
. 3
Operating the Maxi Sky 2 Plus in Dual Mode
19
Spreader Bar Description
WARNING:
Only the spreader bar shown below must be used to perform seated transfers when the
Maxi Sky
2 Plus is in Dual Mode. Doing otherwise may lead to injuries.
Green
The "foot-end" spreader bar is reserved
for the patient's lower body. For seated
transfers, only leg sling loops can be
attached to it.
identifiers: feet symbol
This is the only part of the spreader that
rises or lowers when the
or
buttons are pressed on the hand
control.
The "head-end" spreader bar is reserved for the patient's
upper body. For seated transfers, only shoulder sling loops
should be attached to it.
identifier: head symbol
This part of the spreader does not move when the
or
buttons are pressed on the hand control.
Head-end
Foot-end
Fig. 25
Fig. 26
WARNING:
Scale model 700-19490 shall not be used in dual mode as the accuracy of
measurement of the patient's weight may not be within the tolerances of a Class III scale.
The telescopic link will self-adapt in
length to each spreader bar's position
relative to one other.