001-31249-EN rev
. 3
Technical Specifications
49
g
uidance
and
m
anufacturer
’
s
d
eclaration
e
lectromagnetic
i
mmunity
- f
or
all
e
Quipment
and
s
ystems
The
Maxi Sky
2 Plus is intended for use in electromagnetic environment specified below. The customer or the user of the
Maxi Sky 2 Plus should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment-
guidance
Conducted RF
IEC 61000-4-6
3 V outside ISM bands between
0.15-80 MHz
6 V inside ISM and amateur radio
bands between 0.15-80 MHz
3 V outside ISM bands between
0.15-80 MHz
6 V inside ISM and amateur radio
bands between 0.15-80 MHz
N/A
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
20 V/m
80 MHz to 2.7 GHz
N/A
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz,
1 kHz sine) PM; 18 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz,
1 kHz sine) PM; 18 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
N/A