001-31249-EN rev
. 3
How to Use the Maxi Sky 2 Plus
12
Sleep Mode General Instructions
The ceiling lifts turn their displays off to save energy
whenever they are not being used for more than the
amount of time configured in the parameters (refer to
“Ready state” parameter in the table on page 10).
Therefore, the ceiling lifts need not be turned off after
every use.
The green LED remains solid to indicate that the
battery charge is above 25% or flashes if the ceiling
lift is charging.
The red status LED turns on when the battery charge
drops below 25% of their full charge. The ceiling lift will
beep twice every 10 seconds when the batteries reach
the critical point, that is, less than 5%.
The ceiling lift awakens every 15 minutes to display a
low battery message and goes into sleep mode again
after 30 seconds. Pressing any button on the hand
control will also awaken the ceiling lift
.
Usage Monitoring
The ceiling lifts in the
Maxi Sky
2 Plus are
equipped with a maintenance monitor that informs
the user when maintenance is necessary, based
on the number of transfers, usage history, its
pre-programmed regular schedule or due to
unexpected malfunctions (refer to "LCD Display
Symbols" on page 8).
Required Maintenance
An audible beep and the red LED is used to
indicate that a maintenance or malfunction-related
message is being presented on the display.
Displaying Usage Data
The ceiling lifts in the
Maxi Sky
2 Plus also record
the number of times they have been used to assist
the technician servicing the ceiling lifts.
Accessing Usage Data Using the
Button
Momentary pressing the button will bring the
display into "usage history" screens.
Scroll through the screens using the button.
The ceiling lift will exit the data screens when
pressing the button while the “Average Lifts/
Day” is displayed. It is also possible to exit the
usage history at any time using the or
button.
The following is a list of information available in this
mode:
l
abel
d
etail
1) Total Lifts
Cumulate lifts since initial
start-up
2) Lifts Last Day
Updates at 00:00
3) Lifts Last Week
Sunday to Saturday
4) Average Lifts/Day
Based on last 7 days
Accessing the Usage Data Via the
Monitoring Menu
Pressing and holding the button for 3 seconds
will bring the user into a menu screen which
requires a confirmation to view the data info (see
Fig. 7).
back
monitoring
monitoring
customizing
back
Fig. 7
Navigate through the menu using the or
button, select "monitoring", and press the
button to accept selection.
The following is a list of the information available in
this mode.
1)
Total Lifts
2)
Lifts Last Day
3)
Lifts Last Week
4)
Average Lifts/Day
5)
Hours of Use
6)
Overweight Lifts
7)
Over Duty Cycles
8)
Version / Serial #
NOTE
: Information for items 1 to 4 belong to the
leading ceiling lift and is displayed on both
ceiling lifts simultaneously. On the other hand,
each lift will display items 5 to 8 separately.
To exit any data screen, select “back” then press
to accept.
Emergency Stop with Red Cord
The
Maxi Sky
2 Plus has an easy and rapidly
accessible feature that disables the ceiling lifts
completely in the event that an uncontrolled situation
would happen.
CAUTION:
Do not move the lift by pulling on
either of the red emergency stop cords. This
could cause a malfunction or damage.
CAUTION:
Damage to the ceiling lifts may
occur if their emergency cord is pulled with
excessive force.
When pulling either one of the red cords:
• Both ceiling lifts will stop all activity and their
displays and indicators will shut off as well.
• A small red plunger will come out of the casing.
NOTE: Patients weighing more than 110 kg
(240 lb) might descend slowly when the emergency
stop is activated. The position of the patient (sitting
or reclining) may also change as the patient
descends.