CFS100 SERIES OPERATION MANUAL
SECTION 10: CALIBRATION
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS100 Series Power Source Operation Manual
Page 139 of 146 PN 160951-10
Use the “
Select”
soft key to select the parameter for calibration. If you press the “
Exit
” soft
key from this screen you will be kicked out of the Calibration mode and returned to the set
screen.
10.4.3
Voltage: 150.0V Calibration
Use the “
∧
,
∨
" soft keys to navigate to the Voltage 150.0V parameter and press the “
Select”
soft key.
Follow the prompt message provided on the display, and press the “
Test/Reset
” button to
move into the calibration screen for voltage. If you press the “
Exit
” soft key at this screen
you return to the calibration mode screen.
Enter the voltage reading from the voltmeter with the numeric keypad. When the value has
been selected press the “
Enter
” soft key and you will be moved to the next calibration
parameter Voltage 300.0V. If you press the “
Esc
” soft key you will be returned to the
calibration mode screen.