CFS100 SERIES OPERATION MANUAL
SECTION 7: REMOTE PROGRAMMING
Entire Contents Copyright
2016 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
APS CFS100 Series Power Source Operation Manual
Page 104 of 146
7.5
Parameter Setup Command and Query Syntax
These commands are used to modify individual test parameters within each step. Many of
these commands require a parameter value to be sent with the command. The
corresponding query command will return the parameter setting in effect. Parameter values
are sent as numeric values so no unit info like V or Hz should be appended. Also, when the
query commands are sent, the responses will not include any unit characters.
Command
Description
Value
Unit
AR {0|1|2}
Set Auto Run
0=PROGRAM,1=MANUAL, 2=DC
AR?
Return Auto Run Value
0~2
MC <value>
Set Memory Cycle Value
0~9999 ,0=Contine,1=OFF
MC?
Return Memory Cycle Value
0-9999
MEMORY
<value>
Memory Number
1-50
MEMORY?
Return Memory Number
1-50
STEP <value>
Step Number
1-9
STEP?
Return Step Number
1-9
VOLT <value>
Set Voltage Value
0.0~300.0
V
VOLT?
Return Voltage
0.0~300.0
V
RANG {1|0}
Range Set
0=HIGH,1=AUTO
RANG?
Return Range Set
0-1
FREQ <value>
Input Frequency Value
45.0~500
Hz
FREQ?
Return Frequency Value
45.0~500
Hz
AHI <value>
Set Current High Limit
0.000~42.00
A
AHI?
Return Current High Limit
0.000~42.00
A
ALO <value>
Set Current Low Limit
0.000~42.00
A
ALO?
Return Current Low Limit
0.000~42.00
A
APHI <value>
Set Current Peak High Limit
0.0~59.0
A
APHI?
Return Current Peak High Limit
0.0~59.0
A
APLO <value>
Set Current Peak Low Limit
0.0~59.0
A
APLO?
Return Current Peak Low Limit
0.0~59.0
A
PHI <value>
Set Power High Limit
0.0~5000
W
PHI?
Return Power High Limit
0.0~5000
W
PLO <value>
Set Power Low Limit
0.0~5000
W