7.5.
7.5.
7.5.
7.5.
Other Functions
Other Functions
Other Functions
Other Functions
7.5.1.
7.5.1.
7.5.1.
7.5.1.
Re
Re
Re
Re----Zero
Zero
Zero
Zero Operation
Operation
Operation
Operation
Performing this function, a gross display is zeroed and the current displayed value is
used as a standard point.
The operation of the function can be performed from the front panel key, the input
terminal of the I/O and command input.
The adjustable range is based on the zero calibration and
Zero range
[
genf- 6
] of
the function list. The range is displayed in the unit of percentage of the weighing
capacity.
The re-zero data is stored in memory even without power.
Resetting the re-zero data, use the I/O terminal or the F1 ~ F4 key that is set to
Zero
clear
[
1n f-
nn] [
2
].
nn: terminal number of I/O.
Caution
When the internal A/D converter is out of range, the re-zero operation be not
performed and the error code
ZR.ERR
ZR.ERR
ZR.ERR
ZR.ERR
is displayed.
Concerning Parameters of the Function
Changing the range to do the re-zero operation.
[
genf- 6
]
Zero range
The maximum range is 30% of the capacity.
[Function] - [Function setting] - [General] - [Weighing]
Using the prohibition of the re-zero operation in the unstable condition.
[
genf- 9
]
Tare and zero compensation at unstable status
[Function] - [Function setting] - [General] - [Weighing]
Turning on the display, the re-zero function is used.
[
genf-13
]
Clear mode at power ON
[Function] - [Function setting] - [General] - [Weighing]
7.5.2.
7.5.2.
7.5.2.
7.5.2.
Zero Tracking Function
Zero Tracking Function
Zero Tracking Function
Zero Tracking Function
This function automatically traces the weighing deviation at the nearly zero point and
keeps the zero display of the gross display.
Concerning Parameters of the Function
Changing the time of zero tracking
[
genf- 7
]
Zero tracking time
[Function] - [Function setting] - [General] - [Weighing]
Changing the range of zero tracking
[
genf- 8
]
Zero tracking band width
[Function] - [Function setting] - [General] - [Weighing]