![AND Multi Function Weighing Indicator AD-4402 Instruction Manual Download Page 110](http://html.mh-extra.com/html/and/multi-function-weighing-indicator-ad-4402/multi-function-weighing-indicator-ad-4402_instruction-manual_2941316110.webp)
Page
108
AD-4402
10.4.
10.4.
10.4.
10.4.
Parameter List
Parameter List
Parameter List
Parameter List
Category address: [Function] - [Function setting] - [General] - [Weighing]
Category
address
symbol
Name Descriptions
Range
and
choices
Default
genf- 1
Display refresh rate
1: Five times per second
2: Ten times per second
1 to 2
2
genf- 2
Digital filter
0: Not used
1: 11 Hz
2: 8.0 Hz
3: 5.6 Hz
4: 4.0 Hz
5: 2.8 Hz
6: 2.0 Hz
7: 1.4 Hz
8: 1.0 Hz
9: 0.7 Hz
Select 3dB band for two
low pass filters.
0 to 99
4 8
genf- 3
Sampling
freqeuncy divider
Use to decrease the cut-off frequency of
the digital filter.
0 to 10
1
genf- 4
Stability detection
time
0.0 to
9.9
s
1.0
s
genf- 5
Stability detection
band width
The detection condition concerning
satblility.
0.0 : stable at anytime.
0.0 to
9.9
d
2.0
d
genf- 6
Zero Range
The range to zero the gross display.
Center of range is zero calibration.
Unit: percentage of weighing capacity.
0 to 30 % 5 %
genf- 7
Zero tracking time
0.0 to
9.9
s
0.0
s
genf- 8
Zero tracking band
width
The function automatically traces the
weighing deviation at nearly zero
point and keeps zero display of gross
display.
0.0 to
9.9
d
0.0
d
genf- 9
Tare and zero
compensation at
unstable status
When unstable weghing, whether zero
or tare command is used.
0 :Disabled
1 :Enabled
0 to 1
1
genf-10
Tare at negative
GROSS weight
When negative weghing, whether tare
command is used.
0 :Prohibit tare.
1 : Permission to tare.
0 to 1
1
genf-11
Preset tare
Preset tare of material code
0 :Not used
1 :Use
0 to 1
1
genf-12
Preset tare =0
choice
1: If tare value of material code is zero,
the last tare value is used.
2: If tare value of material code is zero,
Tare is set to zero.
1 to 2
1
First filter
Second filter
A/D converter