7.3.5.
7.3.5.
7.3.5.
7.3.5.
Discharge
Discharge
Discharge
Discharge SSSSequence
equence
equence
equence
The discharge sequence is used to discharge the material from the hopper and clear
the hopper after finishing a batch weighing.
Concerning Parameters of the Function
Storing the time between receiving a start command and opening the discharge gate.
[
5q f-38
]
[Function] - [Function setting] - [Sequence] - [Timer] - [Discharge start
delay timer]
Using the alarm for the discharge time limit.
[
5q f-39
]
If it is over, an error
SQ.ERR 7
is displayed.
[Function] - [Function setting] - [Sequence] - [Timer] - [Discharge
monitoring timer]
Storing the time between cleared hopper and closing the gate.
[
5q f-40
]
[Function] - [Function setting] - [Sequence] - [Timer] - [Discharge gate
close delay timer]
Discharging it automatically when the weighing is finished.
[
5q f-14
]
When the finish signal is turned off, the discharge start timer starts.
[Function] - [Function setting] - [Sequence] - [Control] - [Batch finish
action]
Discharging it automatically when the mixing weighing is finished.
[
5q f-17
]
When the finish signal is turned off, the discharge start timer starts.
[Function] - [Function setting] - [Sequence] - [Control] - [Recipe finish
action]