Maintenance and care
60
Edition 05/2014
PlasmaQuant PQ 9000
7. Place the hose guide over the hose and fix it with the clamping arm.
1
Hose guide
2
Clamping arm
3
Speed controller
4
Pump hose
5
Mixing cup
6
Connection to suction hose for
wash solution
7
Connection to pump hose wash
cup
8
Block for clamping the hoses
9
Pump rolls
Figure24
Automatic sampler pump
8.3.3
Clean-up after cup overflow
If the wash cup has overflowed during an analysis run, immediately interrupt the workflow and
clean the device.
1. Stop the workflow immediately.
2. Take up the liquid with cellulose wadding or a cloth. Wipe the surface dry.
3. Ensure that the outlet can be drained, i.e., remove any sharp bends in the draining hose or
make sure that the draining hose does not dip into the liquid in the waste bottle.
8.3.4
Replacing the fuses
Replace the fuses of the automatic sampler as follows:
1. Switch off the automatic sampler at the power switch and pull the mains cable from the
connection on the automatic sampler.
2. Pull out the fuse holder. To do so, insert a screwdriver blade into the slot in the fuse holder
and carefully pry out the holder.
3. Replace defective mains fuses.
Use only fuses of the type T 5A H 250V, 5 x 20mm.