Installation and commissioning
PlasmaQuant PQ 9000
Edition 05/2014
39
8. Insert the cannula into the holder on the
automatic sampler head. Move the holder
along the Z-axis (down and up) and check
whether the cannula fits through the guide at
the lower end of the head. Fasten the cannula
with the nut on the holder (arrow in figure on
the left).
9. Guide the sample hose initially in a loop
through the eyelet on the cannula holder (1).
10. Thread the hose from the left side through the
eyelet (2) at the lower end of the head.
11. Place the hose at the rear in the eyelets at the
rear of the automatic sampler arm.
12. Connect the hose to the sample hose of the PlasmaQuant PQ 9000.
13. Connect the PlasmaQuant PQ 9000 and the automatic sampler to the interface cable
("Autosampler" on the PlasmaQuant PQ 9000 and "Host" on the automatic sampler.
14. Connect the mains cable first with the mains connector of the automatic sampler and
then with the mains socket.
15. Set the pump speed on the automatic sampler so that the liquid level remains constant
and not too much wash liquid overflows.