Specification
PlasmaQuant PQ 9000
Edition 05/2014
17
Instrument fuses
Fuse number
Type
Protected circuit
S1
10 A NFC 10x38 gG AC 400 V
Spectrometer
S2
6 A NFC 10x38 gG AC 400 V
Tubular heating generator
S3
25 A NFC 10x38 gG AC 400 V
Power supply unit generator
Safety circuits
Monitoring of
Plasma compartment door closure
Torch working position
Cooling
Waste air
Argon supply
Plasma (optical monitoring)
Gas supply
Recommended gas
Inlet pressure
Consumption
Argon 4.8 or superior
Permitted components:
Oxygen
≤ 3 ppm
Nitrogen
≤ 10 ppm
Hydrocarbons
≤ 0.5 ppm
Humidity
≤ 5 ppm
6 bar (5 – 7 bar)
13 L/min
Additive gas (optional)
Oxygen 4.5 or superior
6 bar (5 – 7 bar)
0.02 L/min
Environmental conditions
According to DIN ISO 9022-2:2003/01
Corrosion protection
The device is corrosion-proof against the samples used in the analysis
Operating temperature
+15°C to +35°C, at best +20°C to +25°C
as constant as possible during measuring
Humidity
20 to 90% at 20°C
Storage
Temperature: -40°C to +70°C
Use desiccant
Dimensions and weights
Mass
170 kg
Dimensions (WxDxH)
987 mm x 853 mm x 937 mm
Analytic data
Type of sample
Liquid
Type of nebulizer
Concentric nebulizer
Spray chamber
Cyclone chamber