Installation conditions
PlasmaQuant PQ 9000
Edition 05/2014
23
4.6
Device layout and space requirements
The PlasmaQuant PQ 9000 is a compact device conceived for table-top operation. The space
required is a function of all components needed for the measurement.
The automatic sampler is installed next to the base instrument. Additional space may be
required for a PC and possibly a printer. The PC and printer may also be placed on a separate
table.
The workbench must meet the following requirements:
The dimensions of the workbench for the PlasmaQuant PQ 9000 and the automatic
sampler must be at least 2000 mm x 800 mm. The height should be chosen according to
ergonomic aspects.
The workbench must be capable of bearing a load of at least 300 kg.
The workbench surface must be wipe, scrape and corrosion resistant and must not absorb
moisture.
It must be freely accessible from all sides.
A bottle for receiving non-nebulized sample liquids and automatic sampler wash liquid is
placed underneath the bench.
The recirculating chiller requires a footprint of approx. 400 mm x 600 mm. For unimpeded
cooling air circulation and effective cooling, the surfaces of the housing sides of the recirculating
chiller require a minimum distance of 15 cm from adjacent objects. Due to the dissipated heat
from the recirculating chiller and possible noise it should be placed outside the laboratory. An
extension of the cooling water hoses is permitted if the minimum pressure and the flow volume
are maintained.
Component
Width [mm]
Height [mm]
Depth [mm]
Mass [kg]
On the workbench
PlasmaQuant PQ 9000
990
940
855
170
Automatic sampler
285
510
490
15
Outside the laboratory / (or next to the workbench)
Recirculating chiller
368
576
702
86
Under the work table
Waste bottle
∅
200 mm
400 mm