Installation and commissioning
PlasmaQuant PQ 9000
Edition 05/2014
35
5. Attach the torch to the carriage on the adjusting
rail in the sampling compartment and screw it
on.
Attention!
Tighten the screws firmly so that the gas
supply is connected leak-proof.
6. Fit together the spherical joint connection of the
torch and the spray chamber and secure the
connection with the fork clamp.
7. Push the torch on the adjusting rail upwards
until it engages in the height adjustment.
8. Adjust the torch with the manual height
adjustment (2) so that the tip of the injector in
the torch (1) is situated approx. 1 – 2 mm
below the lowermost winding of the induction
coil.