Specification
14
Edition 01/2014
contrAA 600
Overvoltage category
II according to DIN EN 61010-1
Degree of contamination
2 according to DIN EN 61010-1
Safety class
I
Protection type
IP 20
When connecting further components aside from PC and monitor to the output socket you are in danger of exceeding the
permissible limit value of 5 mA for the drain current.
Instrument fuses
Instrument fuse fittings (5
×
20 mm²) according to IEC 60127.
Fuse number
Type
Protected circuit
F3
T 6.3 A/H
Power socket
F4
T 6.3 A/H
Power socket
F5
T 6.3 A/H
Spectrometer
F6
T 6.3 A/H
Spectrometer
F7
T 3.15 A
Xenon lamp
F8
T 3.15 A
Xenon lamp
Furnace fuse
Type
Protected circuit
TR5-T 100 mA
Graphite tube furnace
Power supply fuse
The power supply fuses may only be changed by service engineers from Analytik Jena AG or by technical personnel author-
ized by Analytik Jena AG.
gL-G instrument fuse fittings (10
×
38 mm²) according to 60947-3.
Fuse number
Type
Protected circuit
F1
32 A/T
Power supply
F2
32 A/T
Power supply
Environmental conditions
according to DIN ISO 90022-2:2003 / 01
Corrosion protection
The device is corrosion-proof for the samples used in the analysis.
Working temperature
+ 10 °C to + 40 °C
Humidity
Max. 93 % at +40 °C
Storage temperature (drying agent) - 40 °C to 70 °C
Dimensions and weights
Mass
160 kg
Dimensions (W x H x D):
785 mm
×
653 mm
×
765 (974) mm
Transport of device
Only possible using the corresponding carrying handles which must be securely
screwed into place