Installation and start-up
contrAA 600
Edition 01/2014
51
attached to each other by encasing and are numbered. The tubes are attached to the
rear of the autosampler using the attachment lug. Marking Wash tube "2".
8. Lead the sample aspiration tube of the HS 60 modular through the hose guide part on
the autosampler arm and attach it on the thin aspirating canula of the autosampler arm.
Note
: The autosampler arm can be moved manually.
9. In the AS-FD feed the dosing tube for the diluent (marking "1") through the tube guide at
the autosampler arm and plug it onto the thicker canula of the autosampler arm.
10. Set the sample tray on the autosampler housing, make sure it clicks into place.
Note
: The control does not start the autosampler, or stop it automatically, if there is no
sample tray.
11. Mount the sample tray.
12. Position the sample cover in such a way that the guiding spout of the autosampler
casing takes hold of the groove of the lid.
13. Switch on the contrAA 600 and start the software.
1
Fluidik module connection
2
AAS connection
3
Suspension mount with adjusting screw
4
Connector for outlet tube
5
Screw for wash tube
Fig. 32
Rear of the autosampler AS-FD