Care and maintenance
66
Edition 01/2014
contrAA 600
23. Screw in the insulating ring by hand and
tighten it moderately using the face span-
ner.
Caution!
Risk of breaking the insulating ring!
Do not jam the face spanner!
24. Check visually that the insulating ring
protrudes approx. 0.3 mm from the jaw.
25. Attach the furnace window to the furnace
jacket and insert the pipetter insert.
26. Insert the knurled pin (arrow) to the stop
into the furnace jaw and the connecting rod.
The connecting rod must be in the front po-
sition.
27. Screw the covering screw to the movable
furnace part.
28. Close the furnace with the
[Close furnace]
button.