Sub MkIIIF
User Manual
Int. Approved
Document ref: ASF3-800-08
November 2018
Page 41 of 79
Copyright © 2009 Analox Ltd. All Rights Reserved.
Commercial in Confidence
ID
LCD Characters
Description
4
3
2
1
24 '9' '9' '9' '9' Display high limit
25 '-' '9' '9' '9' Display low limit
26 'C' 'O' 'N' '8' Configuration zero gain error - used in conversion
calculations
27 'C' 'O' 'N' '5' Configuration conversion divide by zero error
28 'S' 'P' 'I' '1' ADC SPI Fault
29 'S' 'P' 'I' '2' EEPROM SPI Fault
30 'S' 'P' 'I' '3' DISPLAY SPI Fault
31 'S' 'P' 'I' '4' RTC SPI Fault
32 ' ' 'F' 'L' 'O' Flow Fault
33 ' ' 'C' 'N' 'T' Watch dog\stack under\overflow counters
34 'C' 'O' 'N' '6' Pressure selection configuration error
35 ' ' 'C' 'A' 'L' Calibration value passed to OEM sensor is out of range
36 'C' 'O' 'N' '7' EEPROM dbase too large - at configuration time
37 'C' 'O' 'D' 'E' Checksum code error
38 ' ' 'R' 'A' '0' REM RAM fault 0 detected - read\write fault
39 ' ' 'R' 'A' '1' REM RAM array under\overflow error detected
40 'F' 'L' 'S' 'H' REM FLASH fault detected
41 ' ' 'E' 'E' '2' Configuration eeprom checksum error
42 'T' 'E' 'C' 'H' In technician mode
43 'C' 'O' 'N' '9' Error between XML download and manufacturer
configuration
44 'B' 'O' 'O' 'T' Boot loader