Sub MkIIIF
User Manual
Int. Approved
Document ref: ASF3-800-08
November 2018
Page 77 of 79
Copyright © 2009 Analox Ltd. All Rights Reserved.
Commercial in Confidence
15
Specifications
Power Source
External Stabilised 12-30 VDC supply with regulation of better than +/-
300mV.
External 110-230V AC supply (Option only)
Operating
Current
Operator Console : Average 400mA at 24V DC supply
REM1 : Average 130mA at 24V DC supply
REM3 : Average 180mA at 24V DC supply
(not including current drawn by customer’s own injection solenoid on a
REM3
–
injection valves must consume less than 24W
–
ie 1A at 24 VDC)
Fuses
Front Panel Fuses
F1
T1.0A in supply to Operator Console
F2
T315mA in supply to REM1/REM2
F2/3/4 F1.5A in supply to REM3 (permits 1A for injection solenoid)
Display Panel
LCD graphic display, 240 x 128 pixels, with adjustable backlight and
contrast controls
Displays time in hours:minutes:seconds in 5mm high characters
Displays current values of measured parameters using 10mm high large
characters
Alarm Indicators
1 Alarm indicator for gas/environment alarms
1 Fault indicator for communications, calibration and general system faults.
1 Audible Buzzer operating on alarm/fault conditions.
Operator
controls
1 Power Switch to switch instrument on and off
5 Pushbuttons used via Menu System
1 Pushbutton to mute alarms
1 Pushbutton to control display backlight
Oxygen Sensors
Analox oxygen sensor with up to 3 year life in air at NTP (type depends on
range- refer Section 12.1)
Ranges
0 to 600 mBar pO2 (hypobaric), 0 to 1500mBar pO2, 0 to
3000mBar pO2
(enquire regarding alternative ranges)
Accuracy
±1% reading, ±2mBar pO2, ±0.1% reading /°C
Choice of units from mBar, Bar, %, %SEV, kPa, mATA and ATA (define at
time of order, please enquire regarding alternative units)
Pressure Sensor
Analox pressure transducer, with bridge output.
Ranges 0.1 to 2.0 BarA (hypobaric), 0.80 to 10.00 BarA, 0.80 to 36.00
BarA, 0.80 to 60.00 BarA (hyperbaric) (enquire regarding alternative
ranges)
Accuracy
±0.3%FS, ±0.02%FS/°C
Choice of units from BarA, BarG, MSW, FSW, ATA, kPaA, kPaG (define at
time of order, please enquire regarding alternative units)