A full list of messages that may appear is shown below. Note that O2 and Pressure are classed as
non-OEM sensors, whereas CO2, temperature and humidity are classed as OEM sensors.
ID
LCD Characters
Description
4
3
2
1
0 ' ' ' ' ' ' ' ' Blank Message
1 'C' 'O' 'N' '0' No configuration or DCM is currently configuring
2 'C' 'O' 'N' '1' Invalid REM checksum
3 'C' 'O' 'N' '2' Number of sensors exceeded
4 'C' 'O' 'N' '3' No sensors enabled
5 ' ' 'D' 'O' 'G' More than 3 watch dog resets must turn power off
6 'E' 'N' 'B' 'L' Sensor fitted but not enabled
7 'C' 'A' 'L' '0' Calibration Zero fault
–
OEM sensor
8 'C' 'A' 'L' 'S' Calibration Span fault
–
OEM sensor
9 'C' 'A' 'L' 'L' Calibration Low fault - non-OEM sensor
10 'C' 'A' 'L' 'H' Calibration High fault - non-OEM sensor
11 ' ' ' ' 'L' '0' Sensor low value alarm
12 'L' '0' 'L' '0' Sensor very low value alarm
13 ' ' ' ' 'H' 'I' Sensor high value alarm
14 'H' 'I' 'H' 'I' Sensor very high value alarm
15 'b' 'b' 'b' 'b' OEM sensor WARMUP or sensor value with an OEM sensor
comms fault
16 ' ' 'U' '-' 'R' Sensor output under-range
17 ' ' 'O' '-' 'R' Sensor output over-range
18 ' ' 'L' 'C' 'D' Excessive number of parameters selected for display
19 ' ' 'S' 'C' 'F' OEM sensor communications fault
20 'F' 'A' 'I' 'L' OEM sensor fail
21 'N' 'O' 'F' 'L' No GAS flow
–
displayed on appropriate LCD
22 ' ' '3' '2' '0' REM Type
23 'C' 'O' 'N' '4' Too many injection controllers have been configured