32
Amico Clinical Solutions Corp.
Normal System/LCH with Camera
1. Ensure 9-pole slip ring is removed from the light head.
2. Ensure the unit is disconnected from the power supply, with no possible accidental reconnection.
3. Install light head as a normal light head.
4. Fish camera cable through the yoke axle. The cables should now be arranged as shown below.
5. Align connectors of the 9-pole slip ring.
Installation: Amico iCE 30m LED Light
Power Cable
Fish the Power Cable
through this end.