Installation: Electrical
Routing The Supply Lines Through The Ceiling Tube
!
WARNING – Electric Shock Hazard
• If the supply lines, position 2 and position 3 are damaged, the pendant system may be live.
• Carefully route in the supply lines, position 2 and position 3.
Grounding The Flange
!
WARNING – Electric Shock Hazard Due To Defective Insulation:
• If the insulation is defective, the pendant system central axis may be live.
• Always ground the pendant system central axis.
• Strip the 2.5 mm² ground conductor (not supplied) and connect it to the grounding point on the flange.
Connecting The Power Supply
Qualification:
• Electrical connection of the pendant system central axis may only be carried out by a qualified electrician as per
NEC 70.
• Follow the Safety Instructions, pages 10 - 11.
• Stranded supply lines installed by the customer must be fitted with wire end connector sleeves.
• It is the customers responsibility to provide either a switch or breaker to allow for the disconnection of mains
from the power supply.
1. All conductors must be secured to the interface plate using the strain relief.
2. Feed the power supply leads through the strain relief.
3. Connect the power supply lines to the terminal block, as shown in pages 40-42.