Installation: Spring Arms
Removing the Safety Plug
1. Use a 2 mm Allen key to unscrew the M3 screw [3] and take out the Cotter holder cover [3].
2. Rotate the Cotter holder 180° so that the slot in the Cotter holder [1] lines up with the slot in the stem.
The Cotter pin [2] should be exposed.
3. Using a small flat head screwdriver, remove the Cotter pin [2] from the groove in the device end stem.
4. Once the locking key has been removed, slide the safety plug out from the lighthead end of the spring arm.
Item
Description
1
Cotter Holder
2
Cotter Pin
3
Cotter Holder Cover
4
M3 Screw
5
Groove