Electromagnetic Compliance Data for iCE 30m Series
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The equipment or system is intended for use in the electromagnetic environment specified below. The customer or
the user of the equipment or system should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
CISPR 11
Group 1
The equipment or system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
The equipment or system is suitable for use in all establishments other than
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations / Flicker
Emissions
IEC 61000-3-3
Complies