Operation: The Amico iCE LED Light

Lighthead Control Panel

1.  Turning the light On and Off: The LED will light up when the lamp is in the On position.

2.  Adjusting the light brightness: The brightness can be adjusted to 50% to 100% (total of 5 steps).

3.  Switching light to endo mode: To deactivate endo mode you can either press Endo button again or press any 

brightness buttons.

4.  To activate the depth light press the Depth button: To deactivate the depth light press the Depth button again. 

This feature enables an optimum illumination of the wound field according to its texture and shadowing effects. 

This is very important in case of small and deep wound channels.

5.  Use Sync feature to synchronize your settings with the other lights (only applies for multiple lightheads available 

at extra cost).

6.  A blinking red light indicates the failure of one of the fuses or the emergency by-pass switch has been activated.

Wall Control Panel for Light (Only if Purchased)

•  The Wall Control Panel will have the exact same features as the Lighthead Control Panel. The Wall Control Panel 

is a feature provided at an additional cost.

•  There is a two second delay in the Power button of the Wall Control. If the light does not turn on, but the LED 

indicator turns on, you will have to press Wall Control, wait 2 seconds, and turn it back on.

Sync

Depth

Endo

Power

Brightness

Lighthead Control Panel

1

2

6

Press to increase 

brightness

Press to decrease 

brightness

3

5

4

Summary of Contents for ICE 25

Page 1: ...Installation and Operation Instructions iCE LED Surgical Lights iCE 25 iCE 30 w w w a m i c o c o m...

Page 2: ...tallation The Central Axis 17 20 Installing the Ceiling Cover 17 Amico Ceiling Covers 18 Installing the Central Axis Core 19 Installing the Central Axis Arms 20 Installation Electrical 21 22 Labelling...

Page 3: ...ration Operation Amico iCE LED Light 42 45 Lighthead Control Panel 42 Wall Control Panel for Light 42 43 Emergency Shut Down 43 Camera Control Box 43 Emergency Bypass Switch 44 Beam Size Adjustment 44...

Page 4: ...lure to adhere to these instructions may result in damages to equipment or injury to users MARKINGS The iCE 25 and iCE 30 LED Series Surgical Lights are designed to comply with the following Standards...

Page 5: ...www amico com 5 Typical Drawings Single Ceiling Mount Single Ceiling Mount for Low Ceiling Height Dual Ceiling Mount...

Page 6: ...6 Amico Clinical Solutions Corp Dual Ceiling Mount with Monitor Range of Motion for Dual Ceiling Mount Typical Drawings...

Page 7: ...41 Cavity 97 100 100 Single Mask with Cavity 41 39 39 Double Mask with Cavity 40 39 39 Number of LEDs main 80 52 52 Number of Other LEDs 320 208 208 LED Service Life hours 50 000 50 000 50 000 Total R...

Page 8: ...the attaching bolt must be adjusted accordingly For false ceilings the suspension tubes for all lamps can be mounted directly through to the solid ceiling The opening required for this purpose can be...

Page 9: ...s been adjusted to a horizontal position as the arm is under spring tension and can bounce up Do not remove tension screw from the spring arm joint until the lighthead has been securely fastened onto...

Page 10: ...e focusing light handle metal plastic provided with lighthead is used without a disposable cover the handle is not protected by a sterile covering Sterilizable handles are available through Amico Clin...

Page 11: ...tions of optic assembly when cleaning always wear rubber gloves and use only a clean white lint free cloth when wiping surfaces Do not touch the glass portion of the lamp with bare fingers Skin oils c...

Page 12: ...rect position e g drilling of a reinforcement rod the structural engineer responsible must be contacted for re assessment since static loading conditions may have changed endangering the loading distr...

Page 13: ...x M16 Lock Washers 12 x M16 Flat Washers 12 x M16 2 Hex Nuts Single Mount 8 x M16 Threaded Rods 48 x M16 2 Hex Nuts 32 x M16 Lock Washers 32 x M16 Flat Washers 2 Flanged Ceiling Tube 6 x M16 Threaded...

Page 14: ...CRIPTION SHEET OF 1 1 PROJECTION PRIVATE AND CONFIDENTIAL THIS PRINT IS PROPERTY OF AMICO CLINICAL AND IS LOANED IN CONFIDENCE SUBJECT TO RETURN UPON REQUEST AND WITH THE UNDERSTANDING THAT NO COPIES...

Page 15: ...tom of the finished ceiling 4 Attach the associated hex nut lock washers and flat washers in the correct configuration as shown below Tighten the M16 hex nuts to 145 lbs ft using a torque wrench WARNI...

Page 16: ...d the flanged ceiling drop tube 2 Push the flanged ceiling tube onto the six M16 threaded rods of the anchor plate 3 Secure the flanged ceiling tube with two M16 hex nuts NOTE Precise horizontal align...

Page 17: ...ly in place Installing The Ceiling Cover 1 Insert ceiling cover to the flanged ceiling tube 2 Slide the retainer ring right under the ceiling cover and tighten with the set screw to hold it in place I...

Page 18: ...aded rods on the anchor plate so that it drops down enough to the ceiling 2 Slide the cover disk up the tube 3 Install the ceiling covers on the M10 rods using 10 32 screws 4 Install the plastic screw...

Page 19: ...ads in the flanged ceiling tube are 90 offset The central axis can thus be mounted on the flanged ceiling tube at an offset of 90 This can help align the stops in the proper orientation Extra provisio...

Page 20: ...the two M4 screws securing the access cover Once the cover is removed all connections can be easily made WARNING Risk of Pendant System dropping Tighten the M10 1 5 countersunk screws to 7 5 lbs ft u...

Page 21: ...lower extension arm C in Position 1 WARNING Electric Shock Hazard If the extension leads are damaged the pendant system may be live Carefully route the extension leads A through the flanged ceiling tu...

Page 22: ...ive the pendant system central axis may be live Always ground the pendant system central axis Strip the 2 5 mm ground conductor not supplied and connect it to the grounding point on the flange Connect...

Page 23: ...ustrates possible spring arm configurations Operating Modes The pendant system central axis and the mounted spring arms are not intended for operation in potentially explosive atmospheres The pendant...

Page 24: ...alled removed Observe the following instructions 1 Firmly grasp the ends of the covers and pull them apart until they separate from one another and are free from the arm 2 To install the covers line u...

Page 25: ...on arm and lift the slip ring up and out of the way 4 Slide the mounting stem upwards into the extension arm sleeve 6 and securely hold in place 5 Have a second person install the following hardware i...

Page 26: ...t in the nose cone 5 lines up with the slot in the stem The locking key 1 should be exposed 3 Using a small flat head screwdriver remove the locking key 1 from the groove in the device end stem 4 Once...

Page 27: ...ing upwards and may cause serious injury The LED light may separate from the spring arm if the key is not properly in place This may cause serious injury Always make sure that the key is installed pro...

Page 28: ...hen inserting the lighthead Ensure the wiring does not catch or pinch between the spring arm stem and device 7 Insert the locking key 1 8 Turn the nose cone 180 until the locking key door screw slot i...

Page 29: ...king key before installation to ensure smooth movement 4 Fish the camera and power cables from the top of the drop tube until it comes out the end of the spring arm Do not fish the camera controller b...

Page 30: ...the spring arm through the yoke axle 8 Fish camera cable through the yoke axle The cables should now be arranged as shown below 9 Align lemo connectors 10 Connect lemo connector and two pin connector...

Page 31: ...Light Protective Cap of the Light Yoke WARNING Be very careful not to bend the lemo cables 11 Install protective cap of the light yoke with the Allen Key 12 Re install or re tighten the brake screw a...

Page 32: ...era Controller Box to the monitor s Input Installation Amico iCE LED Light 55 East Wilmot St Richmond Hill Ontario L4B 1A4 CANADA Tel 905 764 0800 Fax 905 764 0862 DESCRIPTION VATE AND CONFIDENTIAL CO...

Page 33: ...up the tabs and slots and then press in In order to carry out brake adjustments observe the following instructions 1 Ensure that the adaptor or lighthead has been installed 2 Locate the brake screw on...

Page 34: ...mm Allen Key and pry the first cover off 3 Remove the last two screws from the second main cover then remove the final cover Be sure not to damage the sliding flap covers 4 when removing the main cove...

Page 35: ...s can be pinched or cut if inserted into the load adjustment window of the spring arm body Exercise caution when making adjustments to components inside the spring arm body Components inside will move...

Page 36: ...4 Rotate the nut counter clockwise to bring the lighthead end of the spring arm INWARD towards the rest of the spring arm 5 Rotate the nut clockwise to bring the lighthead end of the spring arm OUTWA...

Page 37: ...ws with a 5 mm Allen key 3 For angles between 0 and 30 adjust the spring arm away from the stops unscrew all of the vertical stop screws and replace them in the correct positions 4 Replace the covers...

Page 38: ...Insert a flat screw driver into the inner slot 2 Push the back of the screw driver inwards 3 Insert wire on the same side 4 Remove screw driver 5 Pull on the wire slightly to ensure that the wire is p...

Page 39: ...VE Note The power supply is always on the opposite side of a lighthead 2 Insert the remaining negative lines into the slots labeled VE 3 Insert the remaining ground lines into the slots labeled GND Po...

Page 40: ...ert the following positive lines into SEGMENT 1 as shown in the image below a Lighthead 1 b Wall Control 1 c Power Supply 1 4 Insert the following positive lines into SEGMENT 2 as shown in the image b...

Page 41: ...dual lightheads with wall control configurations please take note of the numbering of the wall control and camera camera ready units 1 Insert the following positive lines into the terminal blocks labe...

Page 42: ...very important in case of small and deep wound channels 5 Use Sync feature to synchronize your settings with the other lights only applies for multiple lightheads available at extra cost 6 A blinking...

Page 43: ...dication To deactivate auto focus press either of the two Focus buttons 5 Adjusting the brightness camera iris of the camera as required The Auto Brightness feature is initiated automatically on start...

Page 44: ...ower is directed straight to the LED pods providing the maximum LUX of the respective lighthead Note The maximum brightness is only available if all the LED pods are operational Emergency Bypass Switc...

Page 45: ...www amico com 45 Operation The Amico iCE LED Light Expanded beam size...

Page 46: ...ill Ontario L4B 1A4 CANADA Tel 905 764 0800 Fax 905 764 0862 DESCRIPTION PART NO SHEET OF DRAWING NO 1 1 PROJECTION PRIVATE AND CONFIDENTIAL THIS PRINT IS PROPERTY OF AMICO CLINICAL AND IS LOANED IN C...

Page 47: ...greased once a year with acid free grease 5 When adjusting the brakes or the hinges at the ceiling attachment please refer to Installing the Central Axis Arms on page 20 Attention Before dismounting t...

Page 48: ...e wall mylar controller has stopped functioning The wall mylar controller may need to be replaced Lighthead is drifting Tighten up the exposed screws on the yokes or the central axis using a 3 8 flat...

Page 49: ...ure 700 hPa 1060 hPa Transport Storage MIN MAX Temperature 10 C 50 C Relative atmospheric humidity 20 90 Air pressure 700 hPa 1060 hPa References on the Package RH 20 90 P 700hPa 1060hPa 10 C 50 C Tem...

Page 50: ...8 34 40 463 lbs 3 Extension Arms 34 40 46 465 lbs Central Axis 2 Arm 2 Extension Arms 28 34 371 lbs 2 Extension Arms 34 40 373 lbs 2 Extension Arms 40 46 375 lbs Central Axis 1 Arm 1 Extension Arm 28...

Page 51: ...roved for the Connection of the Following Devices The central axis is only approved for the connection of medical end devices such as monitors operating and examination lamps as well as devices for di...

Page 52: ...onment Guidance RF Emissions CISPR 11 Group 1 The equipment or system uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference...

Page 53: ...e s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage Dips Short Interruptions and Voltage Variations on Po...

Page 54: ...in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in th...

Page 55: ...38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the e...

Page 56: ...rge Shipping and installation costs after the first twelve 12 months will be borne by the customer Amico Clinical Solutions Corp will warrant the Suspension Systems for a period of one 1 year for part...

Page 57: ...amico com 85 Fulton Way Richmond Hill Ontario L4B 2N4 Canada Toll Free Tel 1 877 264 2697 Toll Free Fax 1 866 440 4986 Tel 905 763 7778 Fax 905 763 8587 Email info amico com LT INSTALL INSTR ICE LED...

Reviews: