Installation/Removal of the Spring Arm
In order to access the spring arm to carry out load, parallel and vertical adjustments; as well as access for cable routing,
you must be able to remove the plastic covers surrounding the spring arm.
Removal:
1. Remove all four plug covers [1] first by rotating each one counter clockwise until you hear a *click* and removing
them from the spring arm main covers [2].
2. Remove two screws [3] from one main cover using a 2.5 mm Allen Key and pry the first cover off.
3. Remove the last two screws from the second main cover, then remove the final cover. Be sure not to damage the
sliding flap covers [4] when removing the main covers.
Installation:
4. In order to install the plastic covers, carry out the preceding steps in reverse.
5. Do not attempt to install the main covers with the sliding cover flaps installed.
6. Leave the flap covers extended and raise the spring arm to its highest and lowest positions and slide each flap into
its groove when all other plastic covers have been assembled.