Calibration | 5-23
3. Select
Auto Verify
.
The system begins the automatic verify, and notifies you how long the verify gas
will be applied to the sensor from the RCU:
Then the system displays the recovery time remaining:
The system returns to monitoring process readings, and the verify operation is
complete. You can view the results of the last calibration or verification using
Cal/Verify Data
from the
Calibrate Key
menu.
remote Verify
Span Gas in this procedure corresponds to the Verify Gas.
To initiate a remote verify, do the following:
Enter the verify gas value to match the verify gas cylinder you are using to verify
your system and determine the acceptable verify range (
Configure Verify
).
This
value is referred to in this section as a set point value.
1. From the
Calibrate Key
menu, select
Initiate Verify.
2. Select
Continue Verify
to continue or
Abort Verify
to cancel the calibration
and avoid recovery time. Pressing
Cancel
acts the same as
Abort Verify
.
3. Select
Remote Verify
.
The following appears on your display:
3. Press
Enter
once the O
2
reading on the far left of the first display line stabi
-
Span/Verify Gas xx:xx
System Verifying Cal
Recover xx:xx
System Verifying Cal
NOTE
Are the Cal Gas values entered?
O2 xx.x Set xx.x
Press ENTER to Verify
Summary of Contents for TM2000 Pro
Page 14: ...xiv TM2000 Oxygen Analyzer This page intentionally left blank...
Page 20: ...1 6 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 48: ...3 26 TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 88: ...5 16 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 108: ...5 36 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 126: ...7 8 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 142: ...B 4 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 146: ...D 2 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 148: ...E 2 TM2000 Oxygen Analyzer This page intentionally left blank...