Controller / User Interface | 4-21
set Current range option
Defines the range of readings that the 0-to-20 or 4-to-20 mA (milliamp) out
-
put will represent.
You can also choose to set a reverse analog output range, so for example, 0 mA
can correspond to the high end of the oxygen range, and 20 mA can correspond
to the low end of the oxygen range.
The range of readings the current output represents is also a function of the type
of process readings the analog output port tracks.
setting the current range
After selecting an analog output port, set the range of readings that the selected
current output represents as follows:
1. From the
Analog Range Key
menu, select
Set Current Range
.
2. If tracking oxygen, select whether you want to enter the 20 mA value using a
percent or PPM range.
3. Define the new value to correspond to the 20 mA analog output:
where y is the analog output port, and xx is the present value that the current
output represents (for example, 10 % O
2
).
4. If tracking oxygen, select whether you want to enter the 0 or 4 mA value us‑
ing a percent or PPM range.
Define the new value to correspond to the 0 or 4 mA analog output:
where y is the analog output port and xx is the value that the current output cur‑
rently represents.
Out # y 20 ma = xx
New Value?
Out # y 4 ma = xx
New Value?
See the
Select Function
section later in this chapter
for the types of readings the
current output can represent.
Summary of Contents for TM2000 Pro
Page 14: ...xiv TM2000 Oxygen Analyzer This page intentionally left blank...
Page 20: ...1 6 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 48: ...3 26 TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 88: ...5 16 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 108: ...5 36 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 126: ...7 8 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 142: ...B 4 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 146: ...D 2 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 148: ...E 2 TM2000 Oxygen Analyzer This page intentionally left blank...