5-4 | TM2000 Oxygen Analyzer
initiate Cal
Perform an automatic, remote, or manual calibration.
The
Auto Calibrate
and
Remote Calibrate
selections are only available if you
have installed an Auto Calibrate card in the control unit and have a Remote Cali-
bration Unit (RCU). If the control unit detects that these options are not installed, it
will automatically prompt you with a manual calibration operation when you select
the
Initiate Cal
menu option.
Automatic calibrations can also be set to run at predetermined frequencies using
the cycle times defined using the
Set Cal Timers/Set Auto Timers
menu option.
Once you’ve set these cycle times, you don’t have to select the
Initiate Cal
menu
option to run an automatic calibration.
When using a digital input to trigger the auto calibration, make sure that the
Setup Key/Remote Switch Config
selection is set to
calibrate
, and not verify.
If you replaced a cell, or your cell has failed a regular calibration
because of a deterioration, you will need to perform a primary cali
-
bration rather than a regular calibration. See the
Primary Calibra-
tion
section in
Setup Key
Chapter 4 for help.
See Chapter 3 for calibration gas requirements.
If you selected
Two-Gas Cal
on the
Setup Key
menu, skip to the
instruction for performing a two-gas calibration on Page 5-18.
NOTE
NOTE
See the
Remote Calibration
Unit Connections
section
in the Chapter 3 for help on
how to trigger an automatic
calibration using a digital
input.
Summary of Contents for TM2000 Pro
Page 14: ...xiv TM2000 Oxygen Analyzer This page intentionally left blank...
Page 20: ...1 6 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 48: ...3 26 TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 88: ...5 16 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 108: ...5 36 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 126: ...7 8 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 142: ...B 4 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 146: ...D 2 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 148: ...E 2 TM2000 Oxygen Analyzer This page intentionally left blank...