Calibration | 5-21
7. Press
Enter
once you have applied this cal gas.
The Zero gas display appears:
8. Press
Enter
once the zero gas reading on the left of the display has stabilized.
If you specified a recovery time, the remaining recovery time displays:
The system then returns to monitoring process readings.
Inject Zero Gas
ENTER to Continue
CANCEL to Abort
O2 xx.x Set xx.x
Press ENTER to Zero
Recovery xx:xx
System Calibrating
The reading on the far right
of the first display line shows
the set point value you
entered for the O
2
zero gas
using the
Cal Gas Values
menu option. This O
2
zero gas
set point value should match
the cal gas cylinder for the O
2
zero gas.