4-16 | TM2000 Oxygen Analyzer
PPM value. This option won’t appear if Alarm 3 is not set to activate based
on an O
2
reading, but is set to activate on the start of a calibration or verifica‑
tion.
3. You will be prompted to define the Alarm 3 set point value:
4. Enter the Alarm 3 set point value, then press
Enter
(just press
Enter
if you
don’t want to change the current value).
5. Select whether you want to enter the Alarm 4 set point using a percent or
PPM range.
You are prompted to enter the Alarm 4 set point value:
6. Enter the Alarm 4 set point value.
7. Press
Cancel
or
Enter
to exit this menu option.
Alarm Configure
Define whether alarms should trigger based on a high-alarm or a low-alarm
condition.
Alarm 3 is set high or low only if it is set to track oxygen levels (%
O
2
). If Alarm 3 is set to trigger at the start of a calibration or verifi
-
cation, this option does not apply.
If you select
Hydrocarbon Upset
as the
Sensor Config
option from the
Setup
Key
menu, Alarm 4 is fixed as both a high oxygen alarm and the alarm that trig‑
gers during a hydrocarbon upset condition. In this case, you will not be able to
select Alarm 4 as a low oxygen alarm. Alarm 3, however, can still be selected as
a high or low oxygen alarm.
Alarm #3 = xx
New Value?
Alarm #4 = xx
New Value?
NOTE
Summary of Contents for TM2000 Pro
Page 14: ...xiv TM2000 Oxygen Analyzer This page intentionally left blank...
Page 20: ...1 6 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 48: ...3 26 TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 88: ...5 16 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 108: ...5 36 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 126: ...7 8 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 142: ...B 4 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 146: ...D 2 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 148: ...E 2 TM2000 Oxygen Analyzer This page intentionally left blank...