5-20 | TM2000 Oxygen Analyzer
After the recovery period, the system returns to process readings.
Manual Calibrate
To perform a manual calibration, follow these steps:
1. Enter the calibration gas value to match the zero calibration gas you are using
to calibrate your system (
Cal Gas Values
).
2. From the
Calibrate Key
menu, select
Initiate Cal
.
3. Select
Continue Cal
to continue or
Abort Cal
to cancel the calibration and
avoid recovery time. Pressing
Cancel
acts the same as
Abort Cal
.
4. Select
Manual Calibrate
.
Apply the O
2
span gas to the calibration gas inlet port of the sensor (tubing must
be free of oil and dirt):
5. Press
Enter
once you have applied this calibration gas.
The span gas display then appears:
6. Press
Enter
once the O
2
gas reading on the far left of the display has stabi-
lized.
Apply the O
2
zero gas to the calibration inlet port of the sensor:
Recovery xx:xx
System Calibrating
Are the Cal Gas values entered?
Inject Span Gas\Vrfy Gas
ENTER to Continue
CANCEL to Abort
O2 xx.x Set xx.x
Press ENTER to Span
The reading on the far right
of the first display line shows
the set point value you
entered using the
Cal Gas
Values
menu option.
This O
2
zero gas set point
value should match the cal
gas cylinder for the O
2
span
gas.
Summary of Contents for TM2000 Pro
Page 14: ...xiv TM2000 Oxygen Analyzer This page intentionally left blank...
Page 20: ...1 6 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 48: ...3 26 TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 88: ...5 16 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 108: ...5 36 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 126: ...7 8 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 142: ...B 4 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 146: ...D 2 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 148: ...E 2 TM2000 Oxygen Analyzer This page intentionally left blank...