Calibration | 5-15
inject Cal Gas
You must have an auto cal card and RCU installed to use this menu
option.
Manually adjust the pressure and flow of your calibration gas (pressure of cali
-
bration gas cylinders and flow into sensor) by cycling through the solenoid valves
on the RCU. No calibration or verification data is collected.
It is important that all calibration gases, as well as the sample gas,
flow through the analyzer at the same flow rate.
1. Connect the zero gas cylinders to the RCU.
2. From the
Calibrate Key
menu, select
Inject Cal Gas
.
The following message displays, showing that the RCU has introduced the zero
gas to the sensor:
Adjust the pressure and flow for the zero gas to the desired levels. You will be
prompted to end the cycle.
3. Press
Cancel
to end the cycle.
If you specified a recovery time, the remaining recovery time displays.
The system returns to monitoring process readings.
NOTE
NOTE
Zero gas is flowing
CANCEL to quit
Recovery xx:xx
See the
Mechanical Installa-
tion
section in Chapter 3 for
recommended calibration gas
pressure and flow require
-
ments.
Summary of Contents for TM2000 Pro
Page 14: ...xiv TM2000 Oxygen Analyzer This page intentionally left blank...
Page 20: ...1 6 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 48: ...3 26 TM 2000 Oxygen Analyzer This page intentionally left blank...
Page 88: ...5 16 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 108: ...5 36 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 126: ...7 8 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 142: ...B 4 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 146: ...D 2 TM2000 Oxygen Analyzer This page intentionally left blank...
Page 148: ...E 2 TM2000 Oxygen Analyzer This page intentionally left blank...