2-10 | Thermox WDG-V / VC / VCM with Blow Back
Temperature Sensors
There are two tightly controlled temperature zones in the WDG-V. The
Oxygen Cell Temperature and the combustible block temperature (CM
units only). (On oxygen-only units the box temperature is controlled with
an RTD mounted on the aspirator block.)
Cell Temperature
The oxygen cell is controlled to 680 °C. The cell temperature is measured
by dual, Type K thermocouples. Dual thermocouples are used for self
check of the cell temperature. The thermocouples are collocated in a
single assembly. There is no distinction between the two thermocouples
as they are effectively at the same location in the assembly.
Flow Sensor Heater Element (Black Wire)
Flow Sensor Sense Element (Brown Wire)
Combustible/Box Temperature
The combustible block/ box temperature is controlled to 225 °C. This
temperature is measured by 100 Ohm platinum RTD. The RTD is embed-
ded in the combustible block. On oxygen only units, the RTD is mounted
on the aspirator block. All internal plumbing components are maintained
above 200 °C.
Blow Back Thermostat
Located in the blow back enclosure to control temperature to 150 °C.
Heaters
There are three main heaters in the WDG-V, the cell heater the sensor box
heater and the blow back box heater. In CM units, an additional heater is
used for fine temperature control of the combustible block.
Summary of Contents for Thermox WDG-V
Page 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Page 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Page 73: ...AMEVision Display User Interface or Communications 4 5...
Page 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Page 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Page 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Page 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Page 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Page 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Page 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Page 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...