Service and Parts | 6-5
Cell Replacement/Cleaning
Disconnect power from the sensor and control unit and allow the sen-
sor to cool before replacing parts inside the sensor.
Use Figure 8-2:
1. Open the sensor cover to expose sensor components.
2. Remove the cell clips.
3. While holding the top of the cell housing with a backup wrench,
loosen the lower hex nut (not the top hex nut), then remove the entire
cell assembly. (See the
*
in Figure 8-2).
•
Loosening the cell may require a good deal of torque.
•
4. If
cleaning
the cell, wash with water or alcohol. Dry the cell thoroughly
before reinstalling. Always use a new cell O-ring.
• If
replacing
the cell, discard the old cell and cell O-ring and retrieve
the new cell with supplied cell O-ring. Avoid touching the bare cell.
Instead, hold the cell by one of its hex nuts.
5. Place the cell O-ring on the base of the cell as shown in Figure 8-2.
6. Insert the new cell into the cell housing. Do not touch the bare cell
when reinserting. Tighten the cell into the cell housing using the
lower hex nut (the upper hex nut is pre-tightened at the factory). The
cell O-ring provides a seal for the system. Make sure it is aligned into
its proper recess and evenly crushed when tightening the cell.
NOTE
Summary of Contents for Thermox WDG-V
Page 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Page 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Page 73: ...AMEVision Display User Interface or Communications 4 5...
Page 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Page 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Page 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Page 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Page 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Page 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Page 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Page 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...