Troubleshooting | 5-9
Alarm and Warning Messages
ALARMS
ALARM
DESCRIPTION
Possible Causes
Cell T/C Failure
Thermocouple is not connected or is open
Bad T/C, Bad connection
Cold Junction Compensator
Failure
The PCB temperature sensor has failed
Bad Electronics
Cell Temperature Control
A critical overtemperature condition has
occurred
Shorted, open, or Bad
T/C or Failed Temperature
Control Circuit.
Unit must be reset to clear
this alarm
Cell Over Temp
Cell temperature is 4 degrees or higher
above the setpoint
Bad Thermocouple, Failed
temperature control circuit
Cell Under Temp
Cell temperature is 4 degrees or lower
below the setpoint
Faulty heater, faulty heater
connection, Faulty T/C,
Faulty T/C connection,
Low Power.
Only after warm-up is
complete
Cell Temp Rise Failure
Cell heater failed to heat the cell
Bad Cell furnace heater,
Bad heater connection,
Shorted T/C, Bad T/C con
-
nection.
Unit must be reset to clear
this alarm
Over Temp Relay Tripped
The cell over-temperature relay has tripped
Open Thermocouple, Over
Temperature condition,
Failed Temperature control
circuit.
Must cycle power to reset
the relay
RTD Failure
The Box Temp RTD Failed
Open or Shorted RTD
Box Temp High
Box temperature is 4 degrees or higher
above the setpoint
Failed temperature control
circuit
Box Temp Low
Box temperature is 4 degrees or lower
below the setpoint
Faulty heater, faulty heater
connection, Faulty RTD,
Faulty RTD connection,
Low Power.
Only after warm-up time
expires
Box Temp Rise Failure
Box temperature failed to heat
Bad box heater(s), Bad
heater connection, Bad
RTD, Bad RTD connection
Summary of Contents for Thermox WDG-V
Page 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Page 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Page 73: ...AMEVision Display User Interface or Communications 4 5...
Page 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Page 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Page 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Page 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Page 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Page 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Page 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Page 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...