System Overview
|
2-1
SYSTEM OVERVIEW
Sensor Operations
Designed for fast response in a wide range of flue gas applications, the
sensor mounts directly on the combustion process to provide continuous
measurement of oxygen, or oxygen and combustibles, or oxygen, combus-
tibles and methane.
A sample is drawn from the process stream by means of an air-operated
aspirator and is immediately returned to the process. A portion of this
gas rises past the oxygen measuring cell and combustibles detector and
returns to the primary loop. All sample wetted parts are heated to 200 °C
maintain the sample above the dewpoint of the gas.
Observe these cautions when working on the sensor:
Do not turn on the aspirator until the analyzer has been turned on
and is hot - preferably 24 hours. Also, if you turn off the analyzer or
the process is shut down, turn off the aspirator air to avoid plugging
problems!
Remove AC power from the sensor and allow the sensor to cool for at
least one hour before performing any maintenance or troubleshooting
activities.
The outside of the sensor cover and all sensor assembly components
are hot during normal operation (up to 500°F, 260°C inside the cover).
Allow sensor components to cool for at least an hour before working
inside the sensor. Use caution and wear appropriate gloves when han-
dling components or when touching the sensor cover!
•
Summary of Contents for Thermox WDG-V
Page 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Page 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Page 73: ...AMEVision Display User Interface or Communications 4 5...
Page 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Page 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Page 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Page 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Page 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Page 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Page 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Page 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...