1-2 | Thermox WDG-V / VC / VCM with Blow Back
Calibration:
Calibrate or verify calibration. Store last 10 calibration data.
Selectable calibration gas run time and process recovery time.
Timed automatic calibration with optional Remote Calibration
Unit.
Calibration Gas Requirements:
Use calibration gasses @ 10 psig, 3.0 SCFH. Use certified
mixtures only.
O2 Span Gas: Instrument air or from 1.0 to 100% O2, balance
N2 (typically, 2% O2, balance N2).
O2, Combustibles, and CH4 Zero Gas: 2% O2 or from 0.1 to 10%
O2, balance N2.
Combustibles Span Gas (WDG-VC and WDG-VCM only): 60 to
80% of the selected combustibles recorder output range in equal
mixtures of CO + H2, 3-4% O2, balance N2.
CH4 Span Gas (WDG-VM, WDG-VCM only): 2% CH4, 8 to 10%
O2, balance N2.
For example, using a 2000 ppm Combustibles range, span with a
mix of 800 ppm CO, 800 ppm H2, 3-4% O2, balance N2.
System Compliance:
EMC Compliance:
2014/30/EU
Safety Compliance (LVD):
2014/35/EU
IP65 and NEMA 4X
NOTES
1.
All static performance characteristics are with operating variables constant.
2.
System accuracy referenced to 0.1 to 10% calibrated range.
3.
Response is to calibration gas.
4.
Specifications for combustibles and methane apply only to those versions equipped for
measuring these gases.
Summary of Contents for Thermox WDG-V
Page 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Page 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Page 73: ...AMEVision Display User Interface or Communications 4 5...
Page 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Page 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Page 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Page 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Page 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Page 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Page 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Page 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...