2-14 | Thermox WDG-V / VC / VCM with Blow Back
Alarm Contacts
There are 5 alarm contacts in the WDG-V, 1 service alarm, 1 data valid
alarm, and 3 configurable process alarms.
Service Alarm Relay
This relay is used to determine if there is a problem with the analyzer.
This relay is normally open. Contacts are closed when in normal mode.
The relay will be open when there is a problem with the analyzer or on
loss of power.
Data Valid Relay
This relay is used to determine if the output is representative of the
process. This relay is normally open. Contacts are closed when in nor-
mal mode. Contacts are open when there is a problem with the analyzer,
when the analyzer is calibrating, warming up, when the analyzer is in
diagnostic mode, or on loss of power.
Contact Input
There is one Contact Input peripheral on the WDG-V.
An auto calibration can be initiated using the contact input peripheral.
(Only with the RCU option).
Auto calibration must be selected in the sensor configuration menu.
Remote Calibration Unit Interface
The WDG-V has 5 solenoid valve drivers that are compatible with AM-
ETEK’s remote calibration unit (RCU).
See Wiring Section for interface details.
Summary of Contents for Thermox WDG-V
Page 8: ...viii Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 16: ...xvi Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 22: ...1 6 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 34: ...2 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 57: ...Installation Guide 3 17 Figure 3 18 RCU flow diagram 3 Span Combustibles and Methane...
Page 58: ...3 18 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 64: ...3 24 Thermox WDG V VC VCM with Blow Back Figure 3 20 Wiring...
Page 73: ...AMEVision Display User Interface or Communications 4 5...
Page 76: ...4 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 88: ...5 12 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 91: ...Service and Parts 6 3 Figure 8 1a Wiring diagram for sensor...
Page 96: ...6 8 Thermox WDG V VC VCM with Blow Back Figure 8 3 Thermocouple replacement...
Page 100: ...6 12 Thermox WDG V VC VCM with Blow Back Figure 8 5 Flow sensor replacement...
Page 103: ...Service and Parts 6 15 Figure 8 7 Box Heater Replacement...
Page 105: ...Service and Parts 6 17 Figure 8 8 Combustible Block Heater Replacement...
Page 107: ...Service and Parts 6 19 Figure 8 9 Electronics Board Removal...
Page 114: ...7 4 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...
Page 118: ...A 4 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 126: ...A 12 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 132: ...A 18 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 137: ...PC Configurator Software A 23 Figure 23 Calibration_Calibration Results screen...
Page 138: ...A 24 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 144: ...A 30 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 156: ...A 42 Thermox WDG V Series Analyzers This page intentionally left blank...
Page 164: ...A 8 Thermox WDG V VC VCM with Blow Back This page intentionally left blank...