User Manual
AMETEK Programmable Power
RS Series
179
5
FREQ TRANSIENTS
The above test can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.
VOLT MAX
Standard/Group
RTCA
Group1
Group2
Group3
Voltage
1
97
97
104
97
97
3
98.5
98.5
105.5
98.5
98.5
Frequency
400
400
370
360
360
Table 9-10: Abnormal Voltage Minimum
Standard/Group
RTCA
Group1
Group2
Group3
Voltage
1
134
134
122
134
134
3
120.5
132.5
120.5
132.5
132.5
Frequency
400
400
430
650
800
Table 9-11: Abnormal Voltage Maximum
This test will set the voltage and frequency to levels defined by Figure 9-10 for 5 minutes. The
test will be repeated for Group1 only as indicated in Table 9-10 for voltage and Table 9-11 for
frequency. All Groups will repeat the test using Table 9-10 for the voltage setting and Table 9-
10 or Table 9-11for the frequency setting. The
←
key (backspace) will terminate the test at any
time.
VOLT MIN
This test will set the voltage and frequency to levels defined by Table 9-11 for 5 minutes. The test
will be repeated for Group1 only as indicated in Table 9-11. All Groups will repeat the test using
Table 9-11 for the voltage setting and Table 9-10 for the frequency setting. The
←
key
(backspace) will terminate the test at any time.
VOLT UNDER
This test will drop the output voltage from 115 volts to 60 volts for 7 seconds.
VOLT SURGE
This test requires 180 volt output voltage. If the power source is set at the low voltage range,
the high voltage range will be selected before the test starts. At the end of the test the AC
source will be switched back to the low range.
The output voltage will surge to 180 volts for 100 ms. followed by drop to 148 volts for 1 sec
before it returns to 115 volts. The
←
key (backspace) will terminate the test at any time.
FREQUENCY TRANSIENTS (Group 1 only)
Seq. No.
Frequency
Time
1
400
5 minutes
2
480
5 sec.
3
400
10 sec.
4
320
5 sec.
5
400
10 sec.
Table 9-12: Abnormal Frequency Transient
This test will set the voltage at 115V and will remain at this voltage through out the test. The
test will cycle the frequency three times as shown in Table 9-12. Each repeat will start from
sequence 2.
Summary of Contents for RS180-3Pi
Page 1: ...M440036 01 Rev G www programmablepower com RS Series AC and DC Power Source User Manual...
Page 58: ...User Manual AMETEK Programmable Power RS Series 58 Figure 3 14 Functional Test Setup...
Page 133: ...RS Series 133 Figure 5 3 Power Module Detailed Block Diagram...
Page 173: ...User Manual AMETEK Programmable Power RS Series 173 Figure 9 4 Voltage Modulation...