4.2.13 OUTPUT CAL FACTORS Menu
Figure 4-27: OUTPUT CAL FACTORS Menu (Series II only)
The OUTPUT CAL FACTORS menu provides access to the output calibration parameters.
These parameters are password protected and can only be changed after the calibration
password has been entered. For three phase configurations, the PHASE keys toggle between
the three calibration screens for each phase. Refer to the calibration section in this manual for
details on performing a calibration.
The following calibration factors are available from this menu:
Entry
Description
VOLT FS
Full scale voltage output calibration factor.
VOLT ZERO
Zero offset voltage calibration factor.
PHASE OFST
Phase offset calibration factor. Compensates for phase shift
caused by AC amplifier.
IHARM FS
Option –413 Inter harmonic calibration coefficient.
IMP. REAL FS
Not available on RS Series.
IMP. REACT FS
Not available on RS Series.
IMP. REAL MIN
Not available on RS Series.
IMP. REACT MIN
Not available on RS Series.