26
Error
Import/Export: no data found
User information
27
Error
Import/Export: corrupted data
User information
28
Error
Import/Export failed
User information
32
Error
Unknown error from STM
Contact service
33
Error
Error from STM
Contact service
34
Error
Error from STM
Contact service
35
Error
Internal communication error
Contact service
36
Error
Internal communication error
Contact service
58
Error
Fan speed too low, check if fan is functional
Contact service
59
Error
Error in fan parameters, running fan at full speed.
Contact service
60
Error
Temperature sensor not calibrated. Device
temperature may not be valid.
Check cooling/
filters
If error persists,
contact service
61
Error
Cannot start mixer: invalid mixer parameters.
User information
62
Warning
Mixer is not calibrated
Contact service
for calibration
63
Error
Mixer could not reach requested flow
Check that the
inlet pressure is
OK
(2-10 bar)
71
Error
Cannot establish connection to external mixer
User information
72
Error
Data log network dump failed: cannot connect to
server
Check LAN
connection
73
Error
Data log network dump failed: connection to server
was lost
Check LAN
connection
74
Error
Data log network dump failed: no ACK received
Check LAN
connection
75
Error
Data log network dump failed: wrong ACK received
Check LAN
connection
76
Error
Mixer low pressure alarm on gas input X
User information
77
Error
Error from slave mixer
User information
78
Error
Configured flow over maximum possible
User information
80
Error
Cannot start mixer: flow under minimum
User information
81
Error
Pressure alarm on gas input X
User information
No.
Type
Message
User action
Summary of Contents for mocon Dansensor MAP Mix ProVectus
Page 1: ...P N 300862 K 11 2019 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 28: ...26 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 72: ...70 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 79: ...This blank page has been inserted to enable double sided printing of the document...