Dansensor
®
MAP Mix Provectus
EN
User Guide
41
0x0011
Mix 2 setting
(3-gas mixer
only)
1 unsigned byte
Range 0 to 100%
Only used when a 3-gas mixer is connected. In case send to a
2-gas mixer, the mixer will return a ACK, but the received
setting is ignored
For a 3-gas mixer this parameter controls the setting of
"GAS 1 INPUT" i.e. %N
2
See "Mix 1 setting" encoding column, how setting of "Mix 2
setting" will change the setting of "Mix 1 setting"
0x0020
"Mixer table 1"
offset values
(write only)
Not used in MMP
0x0030
"Mixer table 2"
offset values
(write only)
Not used in MMP
0x0040
Vacuum buffer
set pressure
(Write only)
4 byte float. Saved in non volatile memory.
Actual set point for the pressure in the vacuum buffer.
Typical 3.0 or 5.0 bars.
0x0041
Vacuum buffer
hysteresis
(Write only)
4 byte float. Saved in non volatile memory.
Typical 0.3 bar. When buffer pressure reach "Vacuum buffer
set pressure" - "Vacuum buffer hysteresis" the valve is turned
on. When the pressure reaches "Vacuum buffer set pressure"
it is turned off.
0x0042
Vacuum buffer
min on time
(times 100ms)
(Write only)
Not used in MMP, setting is ignored.
0x0043
Vacuum buffer
min off time
(times 100ms)
(Write only)
Not used in MMP, setting is ignored.
Location Register write Encoding
Summary of Contents for mocon Dansensor MAP Mix ProVectus
Page 1: ...P N 300862 K 11 2019 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 28: ...26 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 72: ...70 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 79: ...This blank page has been inserted to enable double sided printing of the document...