42
User Guide
EN
Dansensor
®
MAP Mix Provectus
0x0050
Set mixer type
(Write only)
1 byte unsigned. Saved in non volatile memory.
0 = Not used in MMP (Ignored)
1 = CO
2
/N
2
2 = N
2
/O
2
3 = CO
2
/O
2
4 = N
2
/CO
2
/O
2
NOTE! It is NOT allowed to send 2-gas modes to a 3-gas
mixer or vice versa
0x0051
Set input range
for voltage input.
0-10/2-10V
1 byte unsigned. Saved in non volatile memory
0 = 0 to 10V
1 = 2 to 10V (default)
NOTE! If V-in (voltage input) is not used, always select the
range 2 to 10 V and short circuit the input.
0x0066
Output valve
mode (write)
1 byte unsigned.
See also read 0xAD06
0 = Stop mixer
1 = Start mixer
Location Register write Encoding