12
User Guide
EN
Dansensor
®
MAP Mix Provectus
Flow System
The figure below shows the internal flow system of the
Dansensor®
MAP Mix Provectus
2-
Gas and 3-Gas versions respectively.
The principle used in the gas mixer is based on the use of "Mass Flow Controllers" (MFC). For
each gas line in the mixer there is an MFC controlling the flow of the respective gas. The
mixture is programmed by the software and each MFC is then set to a flow proportional to the
required mixture of the respective gas.
Example
: Total flow required = 500 l/min, 30% CO
2
and 70% N
2
; the software calculates the
necessary flow of each gas input (150l/min CO
2
and 350l/min N
2
) and the MFC´s are set to
deliver these flows.
Each MFC has a PID controller which controls the proportional valve based on the measured
flow from the Mass Flow Sensor and the set point.
The pressure drop through the gas mixer is much lower than on traditional mechanical mixers
and this assures maximum usage of the input gas pressure ranges and also allows running at
lower input pressures (depending on the back pressure at the output).
Each gas line is fitted with a pressure sensor (PS) and in case of low gas supply pressure a gas
input pressure alarm can be set to give an early "Alarm" signal to alert user of low gas supply
pressure.
If input pressure exceeds the minimal or maximal functional levels a "Fault" relay will be
activated. This can be used to stop the packaging machine.
PS
PS
PS
PS
PS
2
2
3
1
1
PS
PS
F3
F2
F2
F1
F1
MFC 1
MFC 1
MFC 2
MFC 2
MFC 3
PV
PV
PV
PV
PV
GAS
IN
GAS
IN
GAS
IN
GAS
IN
GAS
IN
MIX
OUT
MIX
OUT
FLOW/BUFFER
PRESSURE
FLOW/BUFFER
PRESSURE
Proportional
Valve 1
Proportional
Valve 2
Proportional
Valve 1
Proportional
Valve 2
Proportional
Valve 3
Pressure
Sensor 3
Pressure
Sensor 2
Pressure
Sensor 2
Pressure
Sensor 1
Pressure
Sensor 1
Pressure
Sensor 4
Pressure
Sensor 4
Mass Flow
Sensor 3
Mass Flow
Sensor 2
Mass Flow
Sensorr 2
Mass Flow
Sensor 1
Mass Flow
Sensor 1
3-GAS
2-GAS
Summary of Contents for mocon Dansensor MAP Mix ProVectus
Page 1: ...P N 300862 K 11 2019 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 28: ...26 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 72: ...70 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 79: ...This blank page has been inserted to enable double sided printing of the document...