Dansensor
®
MAP Mix Provectus
EN
User Guide
45
Cleaning and Maintenance
General
Performing cleaning and maintenance regularly reduces the chances of equipment failure.
CAUTION! Personnel performing any maintenance or cleaning must familiarize
themselves with the
"Safety Instructions" on page 9
before attempting any of these
procedures.
NOTE! For correct mounting/dismounting of fittings please see
dismounting of fittings" on page 17
Spare parts
For a list of spare parts see
"Spare parts, consumables and accessories" on page 75
.
Cleaning
All of the device’s surfaces should only be cleaned using a mild soap solution and a wrung
cloth.
CAUTION! Never use hard tools or abrasive materials when cleaning any part of the
device.
WARNING! Never use cleaning agents containing chlorinated solvents or acetic or
phosphoric acid. These constitute a health hazard and could damage the
instrument.
Replacing gas inlet filters
Replace the gas inlet filters as described below:
Disconnect the gas inlet hose
and the hose fitting
from the gas inlet filter unit
.
Remove the filter unit
from the device.
Mount the new filter unit
on the device using the gasket
.
Connect the hose fitting
using the gasket
and then connect gas inlet hose
.
3
1
2
5
4
Summary of Contents for mocon Dansensor MAP Mix ProVectus
Page 1: ...P N 300862 K 11 2019 UserGuide...
Page 2: ...This blank page has been inserted to enable double sided printing of the document...
Page 4: ...2 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 8: ...6 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 12: ...10 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 28: ...26 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 72: ...70 User Guide EN Dansensor MAP Mix Provectus P N 300862 K 11 2019...
Page 79: ...This blank page has been inserted to enable double sided printing of the document...