User Guide
Page 5 - 6
Genesis WDG-1200/1210
Insitu Oxygen Probes
(d) Thermocouple fault and heater above maximum limit 40.00.
The analyzer is measuring an unexpected voltage from the probe thermocouple.
CAUSE
ELEMENT
ACTION
1. Thermocouple open circuit.
Probe
Check probe wiring.
2. Thermocouple wired in reverse polarity.
Wiring
Check connections.
3. Thermocouple failed (very unlikely event).
Probe
Replace heater
- thermocouple
assembly.
(e) Cold junction temperature fault 600.0.
The analyzer is reading an unexpected voltage from starting, or correctly
completing, an auto calibration cycle.
CAUSE
ELEMENT
ACTION
1. Sensor not connected or incorrectly wired.
Probe
Check probe wiring.
2. Probe cold junction temperature too high.
Probe
Re-site probe or cool
connection box.
3. Sensor failed.
Probe
Replace sensor
diode.
(f) Failed in calibration 7000.
A fault on the system has prevented the analyzer from starting, or correctly
completing, an automatic calibration.
CAUSE
ELEMENT
ACTION
1. Zero gas pressure not detected.
ACU
Ensure zero cal gas
is connected and
turned on.
2. Any previous fault
See appropriate fault
section.