Genesis WDG-1200/1210
Insitu Oxygen Probes
Span Calibration
The span calibration reading is checked by connecting the calibration gas via
the calibration gas inlet on the probe and pressing
Enter
. The reading displayed
will need approximately 120 seconds to settle.
NOTE:
Recommended span calibration gas concentration is 20.90% O
2
(i.e. ambient air).
Span Calibration Reading
Calibration is necessary if the reading displayed is incorrect (i.e. not equal to
the span calibration gas concentration). It is performed by holding
Enter
for two
seconds. If the reading displayed is correct, the option can be bypassed by
pressing
F1
.