User Guide
Page 4 - 12
Genesis WDG-1200/1210
Insitu Oxygen Probes
Zero setting (AO
↵
ZEro)
To set the minimum O
2
concentration of the
Analog Output
Range
:
Press
Enter
Select a value using the
UP
and
DOWN
keys
Press
Enter
again.
OPTIONS
Set zero reading output
RANGE
0 to 20 %
DEFAULT
0 %
Analog outputs span setting (AO
↵
SPAn)
To set the maximum O
2
concentration of the
Analog Output
Range
:
Press
Enter
Select a value using the
UP
and
DOWN
keys
Press
Enter
again.
OPTIONS
Set span reading output
RANGE
1 % to 25 %
DEFAULT
20.90 %
Analog outputs damping setting (AO
↵
dPng)
To apply a smoothing facility for the output to stabilize the
reading under noisy signal conditions:
Press
Enter
Select a value using the
UP
and
DOWN
keys
Press
Enter
again.
OPTIONS
Setting the damping rate
RANGE
0 to 999.0 sec
DEFAULT
10.00 sec
Base setting (AO
↵
bASE)
The output range required will determine the base selected.
Ranging from 0-10mA to 4-20mA. To make a selection:
Press
Enter
Select the preferred option using the
UP
and
DOWN
keys
Press
Enter
again.
OPTIONS
[0-10][2-10][4-10]
[0-20][2-20][4-20]
RANGE
NOT APPLICABLE
DEFAULT
4-20
Outputs track setting (AO
↵
trAc)
The analog output can be set to “track” the O
2
levels or
“hold” the last reading during calibration. To set track or hold:
Press
Enter
Select
YES
(track) or
NO
(hold) using the
UP
and
DOWN
keys
Press
Enter
again.
.
OPTIONS
[NO][YES]
RANGE
NOT APPLICABLE
DEFAULT
NO
Summary of Contents for LAND Genesis WDG-1200
Page 4: ...User Guide Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Page 6: ...User Guide Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Page 14: ...User Guide Page 2 4 Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Page 68: ...User Guide Page 4 32 Genesis WDG 1200 1210 Insitu Oxygen Probes ...
Page 94: ......
Page 95: ......